즉발감마선 계측시스템의 반사체를 이용한 열중성자 효율증대 연구

Study on Thermal Neutron Efficiency for Neutron Induced Prompt Gamma-ray Spectrometer Using Various Reflectors

분석과학 / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2003, v.16 no.5, pp.426-429
박용준 (한국원자력연구소)
송병철 (한국원자력연구소)
지광용 (한국원자력연구소)

Abstract

Neutron induced prompt gamma-ray spectroscopy (NIPS) system equipped with a $^{252}Cf$ neutron source and a n-type coaxial HPGe detector was installed for the quantitative analysis of aqueous samples in KAERI, Korea. Since the thermal neutron flux for the $^{252}Cf$ neutron source is relatively low compared to that for the reactor, the use of a thermal neutron reflector in the NIPS system may lead to improved results. The enhancement by using various reflectors was carried out by comparing the Cl peak with or without a cadmium plate between sample and the $^{252}Cf$ source. The use of pyrolitic graphite as a reflector provided a good result.

keywords
thermal neutron, reflector, HPGe detector