Determination of lead content in the plastic by ICP and AAS using HI reduction - PbSO4 dissolution method

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2004, v.17 no.5, pp.369-374




Abstract

was decomposed with nitric acid after being carbonized by sulfuric acid. Then excess sulfuric acid and nitric acid was removed by evaporating and drying. Precipitated SO4-2 ion of PbSO4 was resolved itself into H2S by HI in hydrochloric acid and was volatilized. Then Pb+2 was dissolved and was measured by AAS and ICP. The test for reproducibility using PVC-Pb standard material (690-34610mg/kg) was conducted and the accuracy was more than 99.9% compared with the estimated concentration. The precision by AAS was 99.8% and that of ICP was 99.9%.

keywords
Plastics, Lead, Hydroiodic acid reduction, AAS, ICP