- P-ISSN 1225-0163
- E-ISSN 2288-8985
A simple electric circuit of an infrared photodiode electrode (IPDE) was utilized to monitor iron using square-wave (SW) anodic stripping voltammetry (SV) and cyclic voltammetry (CV). The optimum analytical conditions were determined and were compared with those of common working electrodes. The comparison showed that CV is more sensitive and convenient to use than the common voltammetry methods. At the optimized conditions, the working ranges of 0.1- to 0.8- and 0.85- to 6.0 mg/L iron was obtained. Relative standard deviation of 15 measurements of iron (0.4 mg/L) was 0.09%. The analytical detection limit was found to be