    Validation of acid-base titration equivalence between spectroscopic titrator (SPT) and conventional titrators

    Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
    2021, v.34 no.6, pp.267-274
    https://doi.org/10.5806/ast.2021.34.6.267
    Shin, Jiwon
    Kim, Yeajin
    Lee, Hyoseong
    Shin, Jisun
    Yoo, Hyein
    Chae, Won-Seok

    Abstract

    Acid-base titration experiments were carried out using the same sample to obtain a direct comparison of analysis results between a new titrator and conventional Spectrator (SPT) titrators. This allowed the confirmation of an endpoint both by spectroscopic detection and by conventional titrators (automatic burette (ABT) and autotitrator (ATT)). Statistical analysis was performed on the measured values (endpoints) obtained using different titrators, to assess their differences by comparison of straightness, equivalence, ANOVA, F-test, and t-test results. The coefficients of determination, confirmed by straightness, were all >0.999. Thus, the null hypothesis that there was no statistical difference could not be rejected. The Bland-Altman analysis method was used to confirm agreement with the new experimental method. No discrepancy was found in the measured values obtained using different titrators used in this study.

    keywords
    automatic burette, autotitrator, spectrator, statistic analysis, bland-altman plot