Neutron Spectrum Measurement in $n/{\gamma}$ Mixed Field(1)
Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1993, v.6 no.5, pp.501-508
Lee, Kwang-Pill
Kim, Wuon-Shik
Lee,
K., &
Kim,
W.
(1993). Neutron Spectrum Measurement in <TEX>$n/{\gamma}$</TEX> Mixed Field(1). , 6(5), 501-508.
Abstract
In n/${\gamma}$ mixed field of $^{241}Am-Be$(${\alpha}$, n) neutron source, we seperated the neutron component from gamma ray component. At the center of the detector, $^6Li$ was doped on the cerium activated glass plate for $^6Li$(${\alpha}$, n)T nuclear reaction. The time differences of the light following excitations by different scintillators, BC501($C_8H_{10}$) and cerium, and by the same scintillator for different radiations as neutrons and gamma-rays are used to apply the methods of PSD(Pulse Shape Discriminator) and CFD(Constant Fraction Discriminator). The figure of merit of $^6Li$ fast neutron spectrometer is estimated as 1.36.