- P-ISSN 1225-0163
- E-ISSN 2288-8985
Calibration gas mixtures were prepared using dynamic volumetric method according to ISO 6145-5 and the uncertainty was evaluated. Ten identical capillaries with 0.25 mm in inner diameter and 50 cm in length were applied in this system. Dilution ratio of parent gas was determined by the number of capillaries that passes parent gas and that passes balance gas through. Capillaries were made of Teflon which had good chemical stability against adsorption of gaseous substances. Mechanical valves were introduced in this system in order to minimize the thermal effect of solenoid valves. Concentration of prepared gases were compared with master grade standard gases in cylinders made by RiGAS Co. and calibration of the instrument were completed using comparison method according to ISO 6143. Experimental results showed that the coefficient of variance of diluted oxygen standard gases showed less then 0.2% in most dilution range, that of diluted hydrogen sulfide standard gases showed less then 1.0%. Therefore, it is proven that the standard gases prepared by this system are appropriate to be used as a calibration standards in ambient monitoring, etc.