Determination of Cadmium(II) Ion Using the Nafion-Ethylenediamine-Modified Glassy Carbon Electrode
Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2001, v.14 no.2, pp.123-130
Kim, Jin Ah
Ko, Young Chun
Park, Chan Ju
Park, Byung Ho
Chung, Keun Ho
Kim,
J. A., Ko,
Y. C., Park,
C. J., Park,
B. H., &
Chung,
K.
H.
(2001). Determination of Cadmium(II) Ion Using the Nafion-Ethylenediamine-Modified Glassy Carbon Electrode. , 14(2), 123-130.
Abstract
Determination of cadmium(II) ion with a perfluorinated sulfonated polymer-ethylenediamine(nafion-en) modified glassy carbon electrode was studied. It was based on the chemical reactivity of an immobilized layer(nafion-en) to yield complex $[Cd(en)_2]^{2+}$. The reduction peak potential by differential pulse voltammetry(DPV) was observed at $-0.780({\pm}0.005)V$ vs. As/AgCl. The linear calibration curve was obtained in cadmium(II) ion concentration range $5.0{\times}10^{-7}-2.0{\times}10^{-5}M$, and the detection limit(3s) was $2.20{\times}10^{-7}M$. The detection limit of nafion-en modified glassy carbon electrode has been shown about 14 higher sensitivity than a bare glassy carbon electrode.
- keywords
-
nafion (perfluorinated sulfonated polymer)-ethylenediamine modified glassy carbon electrode,
determination of cadmium (II) ion,
differential pulse voltarnmetry